ABTS HRP Microwell Substrate One Component ‘Ready to Use’

ABTS HRP Microwell Substrate One Component ‘Ready to Use’

Product No.: A202

[product_table name="All Top" skus="A202"]

- -
- -
Product Type
Substrate and Assay Reagents
Applications
ELISA
,
Microwell

- -
- -
Select Product Size
- -
- -

Product Details

Storage and Handling
Store at temperatures between 2-8ºC
Country of Origin
USA
Shipping
Next Day Ambient
Expiration Date
Stable for a minimum of 18 months from the manufactured date when stored between 2-8ºC. The substrate should not be frozen and should be protected from direct light by storing in amber bottles. Only high quality amber glass and plastic products should be used for storing aliquots.

Description

Background
ABTS HRPO Microwell Substrate (2.2’-azino-di[3-ethylbenzthiazoline sulfonate (6)]) is a soluble substrate used with the enzyme horseradish peroxidase (HRPO) designed for various qualitative and quantitative immunoassays but not recommended for membrane or immunohistochemical applications where a precipitating reaction product is required. Initially, the substrate should be colorless or very light green in color and is formulated in a mildly acidic buffer. ABTS HRPO Microwell Substrate turns a blue-green color when reacted with horseradish peroxidase labeled conjugates with absorbencies at 405 nm to 410 nm.

Leinco Technologies’ ABTS HRPO Microwell Substrate exhibits superior kinetic performance, sensitivity and lot to lot consistency as compared to other vendors. The outstanding shelf life of at least eighteen months for the ABTS HRPO Microwell Substrate makes this reagent ideal for long term use of the same manufacturing lot.
Directions for Use
ABTS HRPO Microwell Substrate is a ready to use solution that needs no preparation or dilution. Pour estimated amount of substrate into a suitable high quality plastic reservoir to avoid contamination of the bulk solution. It is recommended that you allow the substrate solution to equilibrate to room temperature before use. While the ABTS solution is equilibrating, wash the microplates thoroughly to remove excess peroxidase labeled conjugates. Washing the plates at least four times is recommended to minimize background noise.

Add 100 μl of substrate solution to each well of a 96 well ELISA plate. Once a soluble blue-green reaction product develops, the plate can be read at 405 nm to 410 nm. The absorbance values of the sample should be monitored so that a linear curve can be plotted. The substrate reaction can be stopped by addition of 100 μl of a stop solution such as 1% sodium dodecyl sulfate (SDS) or 405/410 nm Stop Reagent for ABTS Microwell (Leinco Prod. No.: A204). OD values should be read at 405 nm to 410 nm with or without the addition of stop solution. Estimated incubation times for substrate range from 15 to 30 minutes.
Each investigator should determine their own optimal working dilution for specific applications. See directions on lot specific datasheets, as information may periodically change.
Troubleshooting
 
Problem Possible Causes Possible Solutions
High Background Noise Insufficient plate washing · Increase number of washes
· Add detergent or protein to wash solution
· Allow a short soak period between washes
Concentration of enzyme is too high · Check calculations and titrate if necessary
High incubation time  · Reduce incubation time
Contaminated buffers or solutions  · Repeat assay with fresh buffers and solutions
No/Low Signal Capture antibody did not bind to plate · Evaluate coating conditions and standardize
· Increase coating time
 · Increase coating concentration
· Change plate type to high binding
Contaminated buffers or incorrect solutions  · Repeat assay with fresh buffers and solutions
Not enough reporter antibody used  · Increase concentration of HRPO labeled antibody
Poor assay-to-assay reproducibility Inconsistent washing  · Standardize washing and ensure thoroughness
Variations in incubation temperature or time · Ensure constant temperature and incubation time during incubations
· Ensure all reagents are at constant temperature when added
Low reading across entire plate Incorrect wavelength on plate reader · Check filters
· Check absorbance wavelength
Insufficient development time  · Increase development time until background appears

Related Protocols

Elisa Sandwich Protocol
Microwell
Disclaimer AlertProducts are for research use only. Not for use in diagnostic or therapeutic procedures.